R306 and R506 Remote Particle Counters
R306 and R506 Remote Airborne Particle Counters
0.1 CFM (2.83 LPM)
- Industry’s highest concentration of 15,000,000 particles/ft³ @ 10% coincidence loss
- Ideal for use in Cleanrooms, Industrial Health & Safety, and Air Quality applications
- Unparalleled and most comprehensive internal self-diagnostics of any remote particle counter
- Remote diagnostics allows for remote service investigation through the Internet
- Supports MODBUS TCP, RTU & ASCII communication over Ethernet, or (optional) RS-485/RS-232
- Internet of Things (JSON) communication allows for network or cloud based data options
The Airy Technology® R306 and R506 Remote Particle Counters are the most feature-advanced remote particle counters available. Integration is easy into a building automation and cleanroom management system via Ethernet, or (optional) Wireless 802.11 b/g, RS485 or RS232. Integrated web server displays instrument screen through any browser.
The R-306 and R-506 remote particle counters display and report 6 user-selectable particle size channels, as well as (optional) temperature and humidity, flow and sensor status, date of last calibration, serial number, laser on-time and date of manufacture for easy maintenance and warranty management. Storing up to 45,000 time-stamped particle count and environmental data records provides data redundancy on every instrument. Airy Technology® counters can be monitored remotely via web browser or controlled and monitored using Instrument Management
Software (IMS).
All Airy Technology® counters meet ISO 21501-4 and JIS B9921. The R306 and R506 ensure compliance with an on-board pulse height analyzer.
Measures 0.3 μm to 25 μm |
0.1 CFM (2.83 LPM) flow rate |
Long life laser diode technology |
Measures up to 6 channels of simultaneous data |
Approximates mass concentration in μg/m³ |
Large easy-to-use icon driven color touch screen display |
(Optional) temperature and relative humidity probe |
Stores up to 45,000 sample records for on-board data redundancy |
Easy configuration and transferable from instrument to instrument |
Connect using Ethernet, USB or (optional) Wireless 802.11 b/g, RS485, or RS232 |
Static or dynamic IP address (DHCP) |
Internal audible alarm with comprehensive alarm management |
User-selectable channel sizes |
Complies with ISO 21501-4 and JIS B9921 standards |
Easy to clean and wipe down with minimal particle traps |
Versatile mounting options |
Seamless integration into a facility monitoring system |
Light weight stainless steel enclosure |
1 year limited warranty. Extended warranties available. |
Size Range | 0.3 to 25 μm |
Size Channels | Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm |
Counting Efficiency | 50% @ 0.3 μm; 100% for particles >0.45 μm per JIS |
Flow rates | 0.1 CFM (2.83 LPM) |
Concentration Limit | 15,000,000 Particles/ft3 @ 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count | <1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction. |
Count Modes | Automatic, manual, real-time meter, cumulative/differential, count or concentration |
Count Alarms | 1 to 9,999,999 counts |
Calibration | NIST traceable |
Display | 4.3" (10.9 cm) WQVGA (480x272) color touch screen |
Vacuum Requirements | External vacuum >15” (38.1 cm) of Hg |
Number of Channels | 6 |
Custom Size Channels | Calibration for custom size channels available |
Airflow | Internally monitored |
Audible Alarm | Adjustable built-in alarm |
Communication Modes | Ethernet and USB |
Optional Communication Modes | Wireless 802.11 b/g, RS485 or RS232 |
Environmental Sensor | (Optional) temperature and relative humidity probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC), 15-90% ±2% relative humidity |
Alarm | Alarms on counts for all particle sizes, sensor failure, environmental sensors and flow |
Standards | ISO 21501-4 and JIS B9921 |
Calibration | Recommended minimum once per year |
External Surface | Stainless steel |
Dimensions (L x W x H) | 5.2” x 2.25” x 9.96” (13.3 cm x 5.7 cm x 25.3 cm) includes barb fittings |
Weight | 2.6 lb (1.18 kg) |
Accessories | Quick start guide, operating manual on USB flash drive, isokinetic probe, power supply & cable |
Optional Accessories | Printed manual, temperature relative humidity sensor, isokinetic probes, filter, data cable, mounting bracket and sample tubing |
Buffer Memory | 45,000 sample records (rotating buffer) including particle count data and environmental data |
Sample Time | 1 second to 99 hours |
Power | 110 to 240 VAC 50/60 Hz universal in-line power supply |
Operating Conditions | 41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing |
Storage Conditions | 32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing |
Warranty | 1 year limited warranty. Extended warranties available. |
Accessories
Purge Filter
0.1 CFM (2.83 LPM)
Part Number: AS-99002A
Temperature & Humidity Probe
±1° (0.5°C), 15-90% ±2%RH
Part Number: EE-80122A
Remote Mounting Bracket
Part Number: AS-99001
Wireless 802.11 b/g Output
with Antenna
Part Number: EE-80093
Isoprobe Threaded (Ni plated Al)
0.1 CFM (2,83 LPM)
Part Number: PS-12041
Isoprobe Threaded (Stainless)
0.1 CFM (2,83 LPM)
Part Number: PS-12070
Isoprobe Barbed (Stainless)
0.1 CFM (2,83 LPM)
Part Number: PS-12022
Inlet Fitting Barbed
1/8″ (Stainless)
Part Number: PS-12005
Tubing (per foot)
3/8″ OD (1/4″ ID)
Part Number: AS-99019
Tubing (per foot)
1/4″ OD (1/8″ ID)
Part Number: AS-99018
Tubing (per foot)
3/8″ OD (1/4″ ID) Bev-A-Line XX
Part Number: AS-99037
Tubing (per foot)
1/4″ OD (1/8″ ID) Bev-A-Line XX
Part Number: AS-99036
USB Cable
6ft (1.8m)
Part Number: AS-99022
Power Supply
15V -2A 100-240 VAC
Part Number: EE-80127
Power Supply Connector
USA, Japan, Philippines, Taiwan
Part Number: EE-80128
Power Supply Connector
Europe & Korea
Part Number: EE-80129
Power Supply Connector
UK, Malaysia & Singapore
Part Number: EE-80130
Power Supply Connector
China & Australia
Part Number: EE-80131
User Manual
(Printed)
Part Number: MN-24018P
Instrument Management Software
& User Manual (USB Key)
Part Number: MN-24018