Model H8506-30

0.5 – 75 µm @ 0.1 CFM (Large Particles)

The Airy Technology® H8506-30 Handheld Particle Counter is designed for cleanroom monitoring, indoor air quality, industrial hygiene, research, and health & safety applications. Its patented, feature-rich design enables dynamic environmental monitoring and versatile data reporting across multiple industries.

The H8506-30 measures particles from 0.5 to 75 μm in six user-selectable channels at a flow rate of 0.1 CFM (2.83 LPM). It offers data reporting in standard cleanroom formats or by mass concentration in μg/m³, with density and refractive index corrections. Included temperature and relative humidity probes, along with optional CO2 and TVOC sensors, provide comprehensive environmental monitoring.

With unique features like Particles/Second and the Real-Time Meter, the H8506-30 enables efficient, on-the-go particle investigations across nearly any environment. The instrument can also operate as a remote particle counter, leveraging its intelligent battery and power conservation modes to achieve up to 100 hours of battery life.

The Airy Technology® standardized hardware and firmware simplify maintenance and reduce the total cost of ownership. Combined with extensive remote diagnostics and a lifetime warranty, the H8506-30 ensures industry-leading performance and long-term value.

 • Stores up to 45,000 sample records, 1,000 sample locations and 50 recipes
• Long life laser diode technology for reliable performance
• Approximates mass concentration in μg/m³ w/ density and refractive index corrections
• User selectable PM sizes from 0.3 μm to 10 μm including TPM of up to 25.0 μm
• Large, intuitive color touch screen display with an easy-to-use icon-driven interface
• Resolve issues quickly from any location with internet-enabled remote diagnostics
• Annotation function allows user to save 32 character notations to a live sample record
• Over 10 hours continuous sampling on a single charge
• Integrated handle for one-handed operation
• Common interface ensures ease of use across all products
• Connect using Ethernet, USB, Wi-Fi (optional), RS 485 or RS232
• Displays user-definable reports for ISO 14644-1, EU GMP Annex 1 and FS 209E
• Built-in visual and audible alarms with comprehensive alarm management
• Complies with ISO 21501-4 and JIS B9921 standards
• Easy to clean and wipe down with minimal particle traps
• Equipped with a standard 1/4”–20 female mount for tripod compatibility
• Lightweight high-impact injection molded plastic enclosure
• 2 year Limited Warranty (Extended and Lifetime warranties available)
ModelH8506-30
Size Range0.5 to 75.0 μm 
Size ChannelsFactory calibrated at 0.5, 1.0, 2.5, 5.0, 10.0, 30.0 μm variable binning
Flow rate / Accuracy0.1 CFM (2.83 LPM)
Concentration Limits15,000,000 particles/ft3 @ 10% coincidence (per ISO 21501-4)
30,000,000 particles/ft3 @ 10% coincidence (as tested and validated1)
Aerosol Concentration Range0.01 to 20,000 μg/m³
Battery Run Time>10 hours continuous operation
Light SourceLong life laser diode
Counting Efficiency50% @ 0.5 μm; 100% for particles >0.45 μm per JIS
Zero Count<1 count / 60 minutes (<1 particles / 6 ft3). No fault count subtraction.
Count ModesParticles/second, Real-Time Meter, cumulative/differential (count/m³ and count/ft³) and mass concentration (PM)
Count Alarms1 to 9,999,999 counts
CalibrationNIST traceable
Display4.3″ (10.9 cm) WQVGA (480×272) color touch screen
Optional PrinterExternal thermal printer available
Vacuum SourceInternal pump with automatic flow control
Filtered ExhaustInternal HEPA filter
Number of Channels6
Custom Size ChannelsCalibration for custom size channels available
Audible AlarmAdjustable integrated alarm
BatteryRemovable Li-ion (Recharge time: 4 hours within instrument, <2 hours with external battery charger)
ReportsISO 14644-1, EU GMP Annex 1, FS 209E
Recipes50 user-configurable recipes
Communication ModesEthernet, USB, RS485, RS232, and (optional) 802.11b/g wireless
Environmental SensorIncludes temperature (32-122°F, ±1°F / 0-50°C, ±0.5°C) and relative humidity probe (15–90% RH, ±2%) 
AlarmAlarms for particle counts (any size channel), sensor failure, environmental sensors, and flow rate
StandardsISO 21501-4 and JIS B9921
CalibrationRecommended minimum once per year
External SurfaceLightweight high-impact injection molded plastic enclosure
Dimensions (L x W x H)5.12 × 4.25 × 12.26 in (13.0 × 10.8 × 31.1 cm), including handle but excluding probes
Weight2.2 lb (1.0 kg)
AccessoriesUser manual + Instrument Management Software (IMS) on USB flash drive, isokinetic probe, temperature relative humidity sensor, zero-count filter, battery, USB cable, and power supply
Optional AccessoriesPrinted manual, carrying case, spare battery, external battery charger, external printer, barbed fittings, and tubing
Buffer MemorySecurely stores up to 45,000 samples with particle counts, environmental data, locations, and timestamps
Sample Time1 second to 99 hours
Power110 to 240 VAC 50/60 Hz universal in-line power supply
Operating Conditions41-104ºF (5-40ºC) / 20% to 95% non-condensing
Storage Conditions32-122ºF (0-50ºC) / Up to 98% non-condensing
Warranty2 year Limited Warranty (Extended and Lifetime warranties available)